Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Soar 3430: (1 year) CAL VER /5520 DATE: 10-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 36 NUMBER OF LINES: 149 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- R P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: NONE 1.005 DISP Temperature: 18 - 28 degC 1.005 DISP Humidity: <80% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD {DC VOLTAGE} 1.008 DISP Turn the POWER switch to ON. 1.008 DISP 1.008 DISP Press the V button to select DC volts. 1.009 DISP Connect the 5520A to the UUT as follows: 1.009 DISP [32] 5520A NORMAL HI to UUT V/OHM 1.009 DISP [32] 5520A NORMAL LO to UUT COM 1.010 DISP Press the AUTO button to select autorange. 1.011 HEAD DC VOLTAGE: {250mV Range} 1.012 5520 250 200.00mV 0.04% 0.02U 2W 2.001 HEAD DC VOLTAGE: {2500mV Range} 2.002 5520 2500 2000.0mV 0.04% 0.2U 2W 3.001 HEAD DC VOLTAGE: {25V Range} 3.002 5520 25 20.000V 0.04% 0.002U 2W 4.001 HEAD DC VOLTAGE: {250V Range} 4.002 5520 250 200.00V 0.05% 0.02U 2W 5.001 HEAD DC VOLTAGE: {1000V Range} 5.002 5520 1000 1000.0V 0.05% 0.2U 2W 6.001 HEAD {DIODE TEST} 6.002 DISP Press the DIODE button to select diode test. 6.003 5520 500.0mV 1% 0.2U 2W 7.001 HEAD {AC VOLTAGE} 7.002 DISP Press the AC button to select AC volts. 7.003 HEAD AC VOLTAGE: {250mV Range} 7.004 5520 250 200.00mV 0.5% 0.1U 1kH SI 2W 8.001 HEAD AC VOLTAGE: {2500mV Range} 8.002 5520 2500 2000.0mV 0.5% 1U 1kH SI 2W 9.001 5520 2500 2000.0mV 1% 4U 20kH SI 2W 10.001 5520 2500 2000.0mV 5% 20U 80kH SI 2W 11.001 HEAD AC VOLTAGE: {25V Range} 11.002 5520 25 20.000V 0.5% 0.01U 1kH SI 2W 12.001 HEAD AC VOLTAGE: {250V Range} 12.002 5520 250 200.00V 0.5% 0.1U 1kH SI 2W 13.001 HEAD AC VOLTAGE: {750V Range} 13.002 5520 750 700.0V 1% 1U 400H SI 2W 14.001 HEAD {dB} 14.002 5520 -40.00D 1U 1kH SI 2W 15.001 5520 -30.00D 0.15U 10kH SI 2W 16.001 5520 10.00D 0.15U 10kH SI 2W 17.001 5520 10.00D 0.5U 20kH SI 2W 18.001 5520 10.00D 1.2U 50kH SI 2W 19.001 5520 50.00D 0.15U 400H SI 2W 20.001 HEAD {FREQUENCY} 20.002 DISP Press the FREQ button to select frequency. 20.003 5520 10.000H 0.05% 1V SI 2W 21.001 5520 100.00H 0.05% 1V SI 2W 22.001 5520 1.0000kH 0.05% 1V SI 2W 23.001 5520 10.000kH 0.05% 1V SI 2W 24.001 HEAD {RESISTANCE} 24.002 DISP Press the OHMS button to select resistance. 24.002 DISP 24.002 DISP Press the 250 OHM manual RANGE button. 24.003 DISP Connect the 5520A to the UUT as follows: 24.003 DISP [32] 5520A AUX HI to UUT V/OHM 24.003 DISP [32] 5520A AUX LO to UUT COM 24.003 DISP [32] 5520A NORMAL HI to UUT V/OHM 24.003 DISP [32] 5520A NORMAL LO to UUT COM 24.003 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 24.003 DISP the NORMAL leads to the SENSE leads AT the UUT. 24.004 5520 0.00Z S CW 24.005 DISP Press the REL button to zero the displayed reading. 24.005 DISP 24.005 DISP *NOTE* The REL button does NOT work in AUTO RANGE. 24.006 EVAL Does {the reading ZERO correctly (REL TEST)}? 25.001 MESS Press the continuity button (".))"). 25.001 MESS 25.001 MESS Check for an audible tone, and the continuity 25.001 MESS [32] indicator on the display. 25.002 EVAL Does {the CONTINUITY function operate correctly}? 26.001 MESS 26.002 DISP Press the continuity button again to cancel. 26.003 HEAD RESISTANCE: {250 Ohm Range} 26.004 5520 250 190.00Z 0.06% 0.04U CW 27.001 DISP Press the AUTO range button. 27.002 HEAD RESISTANCE: {2500 Ohm Range} 27.003 5520 2500 1900.0Z 0.06% 0.2U CW 28.001 HEAD RESISTANCE: {25 kOhm Range} 28.002 5520 25 19.000kZ 0.06% 0.002U CW 29.001 HEAD RESISTANCE: {250 kOhm Range} 29.002 DISP Connect the 5520A to the UUT as follows: 29.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 29.002 DISP [32] 5520A NORMAL LO to UUT COM 29.003 5520 250 190.00kZ 0.06% 0.02U 2W 30.001 HEAD RESISTANCE: {2500 kOhm Range} 30.002 5520 2500 1900.0kZ 0.5% 0.2U 2W 31.001 HEAD RESISTANCE: {25 MOhm Range} 31.002 5520 25 19.000MZ 2% 0.003U 2W 32.001 HEAD {250mA CURRENT} 32.002 DISP Press the A button to select the current mode. 32.002 DISP 32.002 DISP Press the 250 mA RANGE button. 32.003 DISP Connect the 5520A to the UUT as follows: 32.003 DISP [32] 5520A AUX HI to UUT mA 32.003 DISP [32] 5520A AUX LO to UUT COM 32.004 5520 250 200.00mA 0.25% 0.02U 2W 33.001 DISP Press the AC button to select AC current. 33.002 5520 250 200.00mA 0.75% 0.1U 400H SI 2W 34.001 HEAD {10A CURRENT} 34.002 DISP Press the A button to select the DC current mode. 34.002 DISP 34.002 DISP Press the 10 A RANGE button. 34.003 DISP Connect the 5520A to the UUT as follows: 34.003 DISP [32] 5520A 20A to UUT A 34.003 DISP [32] 5520A AUX LO to UUT COM 34.004 5520 10 5.000A 1% 0.002U HC 2W 35.001 DISP Press the AC button to select the AC current mode. 35.002 5520 10 5.000A 2% 0.01U 100H SI HC 2W 36.001 END